Feature
- For 0.5 to 8 in. Pipes
- No Moving Parts
- 0.05 to 5 m/s (0.15 to 16.4 ft/s) Flow Rate Range
- Accurate Measurement Even in Dirty Liquids
Specifications
Linearity: ±1% rdg ±0.01 m/s (0.033 ft/s)
Repeatability: ±0.5% of rdg @ 25°C (77°F)
Minimum Conductivity: 20 µS/cm
Pipe Size Range: 0.5 to 8″
Flow Range: 0.05 to 5 m/s (0.15 to 16.4 ft/s)Wetted Materials
Sensor Body/Electrodes and Grounding Ring:
PP: Polypropylene/316L SS
O-Rings: FPM (FKM) standard
Other Materials:
Case: PBT
Protection Rating: NEMA 4X (IP65)
Power Requirements
4 to 20 mA: 21.6 to 26.4 Vdc, 22.1 mA max
Frequency: 4.5 to 26.4 Vdc, 15 mA max. Reverse polarity and short circuit protected
Output Specifications
Current Output (4 to 20 mA):
Loop Accuracy: 32 µA max error (25°C @ 24 Vdc)
Isolation: Low voltage <48 Vac/Vdc from electrodes and auxiliary power
Display (Optional):
Characters: 2 x 16
Window: Polyamide
SPDT Relay: 5 A @ 30 Vdc max, 5 A @ 250 AC max
Solid State Relay: 50 mA @ 30 Vdc, 50 mA @ 42 Vac
Maximum Cable: 300 m (1000′)
Error Condition: 22.1 mA
Max Loop Resistance: 300 ?
Frequency Output:
Max Pull-Up Voltage: 30 Vdc
Max Current Sink: 50 mA, current limited
Maximum Cable: 300 m (1000′)
Environmental Requirements
Storage Temperature: -20 to 70°C (-4 to 158°F)
Relative Humidity: 0 to 95% (non-condensing)
Operating Temperature:
Ambient: -10 to 70°C (14 to 158°F)
Media: 0 to 85°C (32 to 185°F)
Max Operating Pressure: 10.3 bar @ 25°C (150 psi @ 77°F), 1.4 bar @ 85°C (20 psi @ 185°F)
Standards and Approvals: CE, NEMA 4X (IP65)