Features
- Two universal sensor inputs, each for RTD, thermocouple, Ω, mV
- Functional safety SIL 2 / SIL 3 in accordance with IEC 61508 (HART variant)
- Various sensor error-calibration possibilities
– One point, two point, Callendar-van-Dusen, freestyle characteristics with 32 supporting points - Sensor backup / automatic sensor redundancy switching (2 x Pt100 3-wire connection and further sensor combinations)
- Redundancy behavior configurable for increased availability, increased safety or increased accuracy (HART variant)
- Output signal
– 4 to 20 mA, HART 5 and HART 7 (switchable)
– PROFIBUS PA 3.01
– FOUNDATION Fieldbus ITK 5.x - Accuracy 0.1 %, long-term stability 0.05 % per year
- Wire break / corrosion monitoring in accordance with NE 89
- Device status signaling and freely configurable diagnostic categorization with diagnostic history according to NE 107 (HART variant)
- Event monitor for logging critical events (HART variant)
- Configuration monitor for configuration changes (HART variant)
- Turnable LCD indicator with operating buttons (option)
- Approvals for explosion protection
– ATEX, IECEx, zone 0
– FM / CSA
– EAC, EAC-Ex
– NEPSI
Applications
- Detection, conversion and transmission of temperature in industrial automation control
- In the field of environmental monitoring, the measurement and transmission of atmospheric temperature, water temperature, ground temperature and other parameters are realized
- Used in weather stations, hydrological stations, geological exploration and other fields
- In the field of medical equipment, it realizes the monitoring and transmission of patient body temperature.
- Applied to the cold chain logistics industry to monitor and record the ambient temperature of logistics vehicles and cold storage
Specification
Output | 4 To 20 MA | Storage Temperature | -50 To 85 °C |
Electrical Connection | Electrical Connection Terminals (captive Screws) Incl. Soldering Tags | Input | Resistance RTD Pt-100, Thermocouples B,E, J, K, N, R,S, T, Voltages, MV Voltages |
Voltage | -125 To 125 MV; -125 To 1100 MV | Ambient Temperature | Standard: -40 To 85°C; Optional: -50 To 85°C |
Electrical Interface | 3.5 KV DC (approx. 2.5 KV AC) 60 S, Input To Output | Vibration Resistance | 10 To 2,000 Hz At 5g In Acc. With IEC 60068-2-6, During Operation And Transport |
Input Resistance | > 10 MΩ | Display | Standard Supply Without Display |
Warm-up Time | 5 Minutes | Power Supply | 11 To 42 VDC |
Housing | Polycarbonate | Communication | HART Protocol |
Calibration | Provided Alongwith And Valid For 1 Year, Traceable To National Standards. | Weight | 50 Grams |