HD32.7 – 8 Inputs Temperature Data Logger – Pt100

Contact us

Features

  • Unit of measurement °C, °F, °K configurable.
  • Flash memory organized in 64 sections with a total capacity of 96.000 acquisitions for each one of the 8 inputs.
  • Storage can be managed in two ways:when the available memory is full, data are overwritten by starting from the oldest ones (circular memory),
  • storage stops when the available memory is full.
  • Simultaneous display of the 8 inputs.
  • Maximum, minimum or average of the stored values.
  • Selectable storage interval: 2, 5, 10, 15, 30 seconds, 1, 2, 5, 10, 15, 20, 30 minutes and 1 hour.
  • Data logging: instantaneous or postponed, with the possibility of selecting the storage start and end.
  • Data download: RS232C, 1200…38400 baud or USB 1.1 – 2.0.
  • DeltaLog9 software for data download and processing.
  • LCD backlit graphic display 128×64 pixel.
  • Instrument setup through the keyboard; no connection required to the PC.
  • Security password for keyboard locking.
  • Power supply: 4 1.5V alkaline C-BABY type batteries or external power supply 12VDC-1A.
  • Consumption @6VDC:
  • <60μA when the instrument is off
  • <60μA in sleep mode with 8 probes connected
  • <40mA during data logging with 8 probes connected

Specification

Temperature Range
Tolerance class WIRE WOUND sensor THIN FILM sensor Tolerance [°C]
classe AA (1/3 DIN) from -50 °C to 250 °C from 0 °C to 150 °C ±(0.1+0.0017∙|t|)
classe A from -100 °C to 450 °C from -30 °C to 300 °C ±(0.15+0.002∙|t|)
classe B from -196 °C to 600 °C from -50 °C to 500 °C ±(0.3+0.005∙|t|)
classe C from -196 °C to 600 °C from -50 °C to 600 °C ±(0.6+0.01∙|t|)

Technical specifications
Measuring Range -200°C…+650°C
Resolution 0.01°C (in the range ±199.99°C) 0.1°C in the remaing range
Internal clock accuracy 1min/month max drift
Unit of measurement °C – °F – K configurable
Memory capacity 96,000 storages for each one of the inputs, max 64 logging session
 

Data Logging

istantaneous or delayed, with the possibility of selecting the storage

start and end time

Storage interval can be selected among 2,5,10,15,30 s;

1,2,5,10,15,20,30 min.;

1 hour

 

Data download

RS232C from 1,200 to 38,400 baud, galvanically isolated. Sub D 9-pole

male connector.

USB 1.1 – 2.0 galvanically isolated.

Security of stored data unlimited
Instrument accuracy when storing ±0.01°C ±1 digit (in the range

±199.99°C)

±0.1°C ±1 digit in the remaining range

 

Power Supply

4 per 1.5V alkaline batteries type C-BABY External 12Vdc-1A power supply.

Connector, external Ø 5.5mm, internal Ø 2.1mm

 

 

Current consumption @6Vdc

<60μA when the instrumen is off

<60μA in sleep mode with 8 probes connected

<40mA during data logging with 8

probes connected

Autonomy 200 hours with 7800mAh alkaline batteries and 8 probes connected
Operating conditions
Operating Temperature -5…50°C
Storage temperature -25 … 65°C
Working relative humidity 0 … 90%RH, no condensation
Protection degree IP64
General characteristics
Dimensions

(Length x Width x Height)

220x180x50mm
Weight 1100 g (complete with batteries)
Materials ABS, polycarbonate and aluminium
Display Backlit graphic LCD 128×64 pixel
Keyboard 15 keys configurable also without PC. Security password for keyboard locking