MTL G1010 Galvanic o2 Gas Analyzer

Eaton MTL G1010 o2 gas analyzer (Crouse-Hinds series G1010 oxygen analysers) is a high-performance oxygen analyzer that can monitor the oxygen concentration in the environment in real time using a primary cell.

  • Electrochemical cell analysers for oxygen
  • Proven technology from gas analyser experts
  • ppm or % O2 levels
  • Hazardous area options
  • Choice of cells for different applications
  • Several Remote or integral sensors options
  • Choice of sample systems

 

Contact us
Category:

An oxygen gas analyzer is a key equipment to ensure the safety of industrial production. Eaton MTL G1010 protects your production with its excellent performance and reliability. If you are interested in this G1010 Gas Analyze, please feel free to contact us!

Features of Eaton o2 Gas Analyzer

Proven Technology from Gas Analyzer Experts
The Eaton electrochemical cell analyzers offer a reliable and accurate solution for measuring oxygen levels in various applications. With decades of experience in gas analysis, we have perfected our technology to provide precise and consistent results.

Measure ppm or % O2 Levels
Whether you need to measure oxygen levels in parts per million (ppm) or percent (% O2), our analyzers can accommodate your specific requirements. Our flexible configurations allow you to choose the appropriate range for your application.

Hazardous Area Options
We understand the importance of safety in hazardous environments. Our analyzers have certifications and features that meet the stringent requirements of hazardous area regulations, ensuring safe operation in challenging conditions.

Choice of Cells for Different Applications
Depending on your application, we offer a variety of electrochemical cells to suit your needs. Our cells are designed to provide optimal performance and durability in different environments, ensuring accurate measurements across a wide range of applications.

Remote or Integral Sensor
Choose between a remote sensor for applications where the analyzer needs to be located away from the measurement point, or an integral sensor for a more compact and integrated solution. Our flexible design options allow you to select the best configuration for your specific needs.

Remote sensor option for G1010 o2 gas analyser

Choice of Sample Systems
We offer a range of sample systems to accommodate different gas flow rates and pressures. Whether you need a simple sampling line or a more complex system with filters and regulators, we have the solution for you.

Excellent accuracy and stability
The use of advanced electrochemical sensor technology ensures the accuracy and stability of the measurement results, and can maintain high accuracy even in harsh industrial environments. The wide measurement range can meet the needs of different industries for oxygen concentration measurement.

Multiple alarm modes
This oxygen gas analyser supports multiple alarm modes such as sound and light alarms, relay outputs, etc., to ensure timely reminders to operators in dangerous situations. Two alarm channels, user-programmable to high, low, or off, each channel provides a set of voltage-free conversion contacts, which can be set to any concentration within the instrument span.

Fast response
Fast response speed, can detect changes in oxygen concentration in time, and provide real-time protection for your production.

Rugged and durable
The rugged industrial-grade design is adopted, which can adapt to various harsh industrial environments and has high reliability.

Compliance with international standards
The o2 analyser complies with relevant international safety standards to ensure safe use.

Multi-function display
LCD display screen, real-time display of oxygen concentration, alarm status and other information.

Self-diagnosis function
With self-diagnosis function, it can detect faults in time and ensure the normal operation of the instrument.

Remote monitoring
The g1010 oxygen analyzer supports remote monitoring, allowing users to understand the equipment operating status at any time.

 

Specification of Eaton o2 Gas Analyzer

Parameter Specification
Stability <2% reading/month or 1 ppm, whichever is greater
Sample Flow 100 – 300 ml/min (optimum), 1000 ml/min (maximum)
Sample Temperature Range –5°C to +40°C, RH <100% (non-condensing)
Sample Pressure Atmospheric (determined by vent pressure) for quoted accuracy
Sample Connections Inlet and outlet: ¼” bulkhead compression fittings, suitable for 0.25 inch (6 mm) o.d. tube
Speed of Response T90: Typically 3 s at % levels, 20 s at ppm levels (varies by sensor and concentration)
Analog Outputs 0–5V (min load 10 kΩ) or 4–20 mA (max load 300 Ω); Isolation: max 30 Vac, 70 Vdc
G1010/G1010R User programmable
G1010Tx/G1010TxX Fixed (refer to manual for full details)
Alarm Outputs 2 alarms, user programmable; Volt-free, crossover contacts
Ambient Operating Range 0°C to +40°C, RH <80% (non-condensing)
Power 110–120V or 220–240V AC, 50/60 Hz, or 24V DC
Enclosure Details G1010: Glass fibre reinforced, Noryl panel mount
Net Weight 1.0 kg
Dimensions (mm) 96 (W) x 96 (H) x 196 (D)

 

Applications of Eaton o2 Gas Analyzer

According to industry, it is widely used in the following industries:

  • Petrochemical industry: Monitor oxygen concentration in storage tanks and pipelines to prevent fire and explosion accidents.
  • Power industry: Monitor dissolved oxygen in transformer oil to prevent transformer failure.
  • Metallurgical industry: Monitor oxygen concentration in high-temperature furnaces to optimize the combustion process.
  • Food industry: Monitor oxygen concentration in packaging gas to ensure food quality.
  • Pharmaceutical industry: Monitor oxygen concentration in the production process of drugs to ensure product quality.

 

According to the specific application, it is used in Nitrogen generators, Inert gas systems, Air separation, Furnace atmospheres, Biogas, Gloveboxes, Gasification, and Packaging.