Features of A40B Series Fluke Current Shunt
- Simplified calibration/verification of precision calibrators and current sources
- Shunt sizes available for currents ranging from 1mA to 100A
- Usable across the dc to 100 kHz frequency range
- Includes 14 individual dc and ac current shunts covering a 1, 2, 5 sequence over 6 decades of current
- Allows for simple direct measurements, eliminating the need for ac/dc transfers
- Demonstrates stability typically better than ± 5.0 µΩ/Ω for one year
- Offers a typical angular accuracy better than ± 0.003º at 1 kHz
Applications of A40B Series Fluke Current Shunt
- Commonly used for controlling large air conditioning systems.
- Used in the control of lighting systems.
- Commonly utilized for compressor control in refrigeration systems.
- Enables normal start-stop and reverse operation in motor control.
- Used to control the connection and disconnection of capacitors in power capacitor banks.
- Applied in the automation control of industrial production lines.
Specification of A40B Series Fluke Current Shunt
Basic Electrical Specification Summary
Shunt
Nominal
Current |
Nominal
Resistance
(Ohms) |
Basic Specification ±μA/A, TCal ±1 °C, ≤50 % RH |
DC |
1 kHz |
10 kHz |
30 kHz |
100 kHz |
1 mA |
800 |
20 |
55 |
75 |
75 |
150 |
10 mA |
80 |
20 |
26 |
26 |
26 |
26 |
20 mA |
40 |
20 |
26 |
26 |
26 |
26 |
50 mA |
16 |
20 |
23 |
23 |
23 |
23 |
100 mA |
8 |
20 |
24 |
24 |
24 |
24 |
200 mA |
4 |
20 |
26 |
26 |
26 |
26 |
500 mA |
1.6 |
21 |
27 |
27 |
27 |
28 |
1 A |
0.8 |
21 |
27 |
28 |
28 |
31 |
2 A |
0.4 |
21 |
27 |
30 |
30 |
48 |
5 A |
0.16 |
21 |
31 |
32 |
40 |
71 |
10 A |
0.08 |
26 |
37 |
60 |
61 |
92 |
20 A |
0.04 |
26 |
43 |
52 |
70 |
113 |
50 A |
0.016 |
32 |
55 |
80 |
81 |
144 |
100 A |
0.008 |
35 |
65 |
90 |
98 |
174 |
Maximum AC-DC Difference
Shunt Nominal Current: |
Maximum AC-DC Difference (± ppm)1,2 |
|
1 kHz |
100 kHz |
1 mA3 |
53 |
150 |
10 mA |
20 |
40 |
20 mA |
18 |
30 |
50 mA |
13 |
16 |
100 mA |
14 |
27 |
200 mA |
17 |
28 |
500 mA |
17 |
21 |
1 A |
17 |
23 |
2 A |
17 |
44 |
5 A |
23 |
69 |
10 A |
28 |
98 |
20 A |
37 |
150 |
50 A |
47 |
180 |
100 A |
60 |
300 |
1 Specifications indicate the maximum flatness deviation from dc, and include both measured AC-DC difference and the uncertainty of measurement. They are stated at k=2, approximately 95 % confidence.
2 Includes 1-year stability of the AC-DC difference.
3 Specifications for the 1 mA current shunt are for TCal ± 1 °C.
General Specifications |
Calibration Documentation |
17025 accredited report of calibration included |
Dimensions (maximum) |
Shunt Value: |
1 mA to 2 A |
Height: |
70 mm (2.75 in) |
Width: |
70 mm (2.75 in) |
Overall Length1: |
124 mm (4.9 in) |
Shunt Value: |
5 A to 20 A |
Height: |
130 mm (5 in) |
Width: |
130 mm (5 in) |
Overall Length1: |
210 mm (8.25 in)) |
Shunt Value: |
50 A and 100 A |
Height: |
200 mm (7.9 in) |
Width: |
200 mm (7.9 in) |
Overall Length1: |
343 mm (13.5 in)) |
1 Includes input and output connectors; subject to change by component vendor. |
|
Physical/mechanical parameters |
Shunt Value: |
1 mA to 20 A |
Weight: |
0.7 kg (1.6 lb) |
Input Connector: |
Type-N (female) |
Output Connector: |
Type-N (female) |
Shunt Value: |
50 A and 100 A |
Weight: |
3.4 kg (7.5 lb) |
Input Connector: |
Type-LC (female) |
Output Connector: |
Type-N (female) |
|
Operating environment |
Temperature: |
13 °C to 33 °C |
Calibration temperature (TCal) range: |
18 °C to 28 °C |
Humidity range for best specification 1,2: |
≤ 50 % RH |
Altitude: |
0 m to 3,000 m |
1 Resistance stability is affected by humidity, but changes are reversible.
2 If the shunts are calibrated outside this RH, stability specifications will be met as long as the shunts are stored and used at the same relative humidity ± 10 % RH. |
|
Storage and transit environments (for models other than the 1 mA current shunt) |
Temperature to avoid damage: |
-20 °C to 140 °C |
Temperature and humidity to maintain performance1: |
5 °C to 45 °C; 15 % to 80 % RH |
Non-operating altitude: |
0 m to 12,000 m |
1 Storage at extremes of temperature or humidity will cause a temporary change of shunt resistance by up to ± 20 ppm. When subsequently stored or used within the limits of the operating environment, the shunts will recover to their original resistance value within 30 days. |
|
Additional 1 mA current shunt specifications |
Output resistance |
8 mΩ |
Maximum safe output current |
11 mA (e.g., 1 V output into 90 Ω) |
Maximum capacitive load |
800 pF |
Output voltage regulation |
15 ppm/100 pF |
Maximum output dc v offset |
± 100 μV (typical ± 25 μV) |
Typical error @ 1 MHz |
< 2 % |
Battery specifications |
Battery size |
AAA (44.5 mm x 10.5 mm) |
Battery technology |
Nickel-Metal Hydride (NiMH) |
Number of batteries required |
8 (in 2 groups of 4) |
Nominal battery voltage |
1.2 V (4.8 V per group of 4) |
Typical battery capacity |
800 mAh |
Storage and transit environment to preserve the batteries |
Less than 90 days: |
-20 °C to 40 °C |
Less than one year: |
-20 °C to 30 °C |
Charging time (from fully discharged): |
100 minutes |
|
Maximum operating time between charges |
Maximum output load (11 mA): |
18 hours |
High impedance load: |
24 hours |
Recommended cooling period: |
100 minutes |
To prevent loss of battery capacity recharge at least twice per year. |
|
|